Product Description
Cleaning and disinfection: after each use, disinfected 5 LED lamps dental high speed handpiece under high temperature and pressure according to following method:
1.Remove the dental diamond bur from the dental handpiece, brush the dirt from handpiece ,wipe it with sterilizing alcohol cotton ball.
2. Lubricated the 5 LED lamps dental high speed handpiece,to wrap and mark with a sterilizing bag
3. Sterilization at high temperature: sterilizing time for 20 minutes under 121°C, sterilizing time for 15 minutes under 135°C
4. Make sure the handpiece is in good working condition when steam disinfected repeated, weekly maintenance is recommended.
5. During repeated steam disinfection of mobile phones, ensure that the phone is in good working condition for regular maintenance. Weekly maintenance is recommended
Note:
1. oil must be injected into the center of shaft from the inlet of diamond bur each day to maintain the service life of the shaft
2.After daily use, spray with a spray lubricant for at least 1-2 seconds each time.
3.If the workload is large, each spray should be taken at noon and in the afternoon.
4.Spray one time use spray lubricant before high temperature disinfection
5.Oil injection lubrication method: take off the dental handpiece, let the fuel spray nozzle point at the inlet hole of the dental handpiece, press it then injection, when injection, the oil storage tank must be upright upward
Storage: stored in a room with a relative humidity less than 80% without corrosive gas and well ventilated.
FAQ: Q: is it 4 holes dental handpiece?
A: Yes, it is. We also have 2 holes, you can choose according to your requirement.
Q: Is it with needle remover?
A: the push button type dental handpiece doesn't need needle remover.